6282 Display Pressure Transmitter
6282 Display Pressure transmitter model no [VAC- 6282-display-pressure-transmitter]
6282_U_500: 0-500
6282_B_250: ± 250
6282_B_500: ± 500 Pa
6282_B_1250: ± 1250 Pa
6282_U_2500: 0-2500 Pa
Non linearity: ± 0,3Pa + 0,1% of reading, typical.
Stability: ± 0,2 % typ.(1 year) Typical
Time constant: 0.1 – 20 seconds
Output: 0-5, 1-5, 0-10, 1-10, 2-10 V 4-20mA, 0-20 mA
Output Range: Unit depending, select in steps
Resolution: 10 bit digital output
Supply: 18-24 V AC/DC, 50 mA
Temp range operation*: -10…50 °C (-14…122 °F)*
Humidity range: 0…95 % RH (non condensing)
Dimension: 125x90x60 mm
Weight: 530 g
VSKU: 1008000037
6282 Display Pressure transmitter Is a top quality digital Processor
The 6282 Display Pressure transmitter is advanced processor or the device that is a transmitter which which is visible through the cover. This aimed for the standard pressure measurement tasks. The device is designed with the user and the installer in mind.
The device reduces many of the problems that traditional and old sensors suffer from and provide as much flexibility as possible.
The processor has a field selectable feature like signal output (mA or Vold) and a Signal volt output (0-10, 0-5, 2-10).
There are different predefined measuring limits i.e. 5, 10 … depending on the selected models are also available. These can be changed by user to other values. For example if you select 25 range and wish to change it to range 0-30.
With these given options you can cover most situations for room pressure measurements in one sensor. For the application with the user selectable rooms modes i.e. switch for positive room pressure, Negative room pressure and neutral room pressure a bidirectional[B] model is used. In a dual room mode the CPS works and handles all 4 alarm i.e. negative high, low, low alarm, and positive high that the 6282 have. The installation is very easy and simple with the 6282 by using the display and push buttons configuring the settings.
The installation is easy and simple with the 6282 using the display and push buttons configuring the settings.
The sensor consists of a silicone membrane and it works with a capacitive technique. The sensor is very accurate and it is durable. For further improving the sensor, the display is visible through the top cover lid so that actual value is readable from far distance (illuminated LED display). Auto calibration is an option that is added on this sensor for applications for the further improvisation where any small drift over the time is not applicable and is not accepted. For example as in the hard to reach locations, or when the sensor is configured for very low pressure like ± 5 Pa.
A simple alarm function can be achieved with the optional relay. For example measuring pressure in room. If the pressure is low a LED light tower is connected directly to the 6282 relay will warn the staff with a light by the door. It is very easy in installations. If wanted Modbus can reach the alarm level. All settings and parameters can be reached from a BMS/Scada through Modbus.
Usages and Applications
It can be used in Room pressure.
It can be used in Laboratories.
It can be used in Static pressure.
It can be used in VAV Boxes.
It can be used in Clean rooms.
It can be used in Hospitals.
It can be used in Isolation room.
It can be used in Industry.
It can be used in IVF clinics.
It can be used in Airlock.
It can be used in Ante rooms.
It can be used in Server rooms.
Specifications
It consists of 0-5, 1-5, 0-10, 1-10, 2-10 V 4-20mA, 0-20 mA output.
The device resolution is 10 bit digital output.
It can supply 18-24 V AC/DC, 50 mA.
The dimension of the device is 125x90x60 mm.
The weight of the sensor is 530g.
It has stability of ± 0,2 % typ.(1 year) Typical.
It consists of ± 0,3Pa + 0,1% of reading, typical.
It takes a time constant of 0.1-20 seconds.